CytoCare Compounded Safety Checks
ITEMS
SAFETY CHECKS
GOALS
IV Bags
individual bag labeling with a unique barcode
assure unique association IV bag / patient
IV Bags
product code scanning just before loading
prevent unintentional exchange of NaCl and Glucose
IV Bags
individual label scanning just after loading
verify the identity of the IV bag when loaded on the carousel
IV Bags
weighing before excess liquid pull-out
measure the real liquid content of the bag
IV Bags
weighing before compounding
verify the actual quantity of liquid taken out, set the tare of the IV bag before drug injection
IV Bags
weighing after compounding
verify the actual quantity of injected drug (0.001 ml)
IV Bags
individual label scanning before unloading
verify the identity of the IV bag when unloaded from the carousel
IV Bags
individual label scanning after unloading, final label is printed
uniquely identify the IV bag before administration
IV Bags
barcode reader and/or visual check of labels consistency
pharmacist, technician, nurse, can check the consistency of identification labels at any moment in the delivery path
Drugs
position check at loading on carousel
verify the position of each drug vial when loaded on the carousel
Drugs
size check before being taken by the robot hand
verify that the vial size is within the expected range
Drugs
computed vision label recognition
verify that the drug to be compounded is the right one
Drugs
weighing before compounding
verify the initial content of the drug vial before compounding
Drugs
weighing after compounding
verify the final content of the drug vial after compounding, verify the actual quantity of injected drug (0.001 ml)
Syringes
position check at loading on carousel
verify the position of syringe when loaded on the carousel
Syringes
position check at loading on syringe holder
verify the position of syringe in the dosing system
Syringes
detection of needle tip position (computed vision)
verify the actual position of the needle tip in space (compensate misalignments)
Syringes
real time laser monitoring of syringe plunger position
verify the actual position of the syringe plunger by an independent system
Syringes
real time monitoring of plunger transport system
verify the actual position of the syringe plunger by a second independent system
© 2007 PRIMUS INNOVATIONS, INC. All rights reserved