CytoCare Compounded Safety Checks

ITEMS SAFETY CHECKS GOALS
IV Bags individual bag labeling with a unique barcode assure unique association IV bag / patient
IV Bags product code scanning just before loading prevent unintentional exchange of NaCl and Glucose
IV Bags individual label scanning just after loading verify the identity of the IV bag when loaded on the carousel
IV Bags weighing before excess liquid pull-out measure the real liquid content of the bag
IV Bags weighing before compounding verify the actual quantity of liquid taken out, set the tare of the IV bag before drug injection
IV Bags weighing after compounding verify the actual quantity of injected drug (0.001 ml)
IV Bags individual label scanning before unloading verify the identity of the IV bag when unloaded from the carousel
IV Bags individual label scanning after unloading, final label is printed
uniquely identify the IV bag before administration
IV Bags barcode reader and/or visual check of labels consistency
pharmacist, technician, nurse, can check the consistency of identification labels at any moment in the delivery path
 
Drugs position check at loading on carousel verify the position of each drug vial when loaded on the carousel
Drugs size check before being taken by the robot hand verify that the vial size is within the expected range
Drugs computed vision label recognition verify that the drug to be compounded is the right one
Drugs weighing before compounding verify the initial content of the drug vial before compounding
Drugs weighing after compounding verify the final content of the drug vial after compounding, verify the actual quantity of injected drug (0.001 ml)
 
Syringes position check at loading on carousel verify the position of syringe when loaded on the carousel
Syringes position check at loading on syringe holder verify the position of syringe in the dosing system
Syringes detection of needle tip position (computed vision) verify the actual position of the needle tip in space (compensate misalignments)
Syringes real time laser monitoring of syringe plunger position verify the actual position of the syringe plunger by an independent system
Syringes real time monitoring of plunger transport system verify the actual position of the syringe plunger by a second independent system
 