Frequently Asked Questions - Installation

What are the dimensions of CytoCare?
CytoCare™ ships from our factory fully assembled with dimensions that are approximately 7 feet of width, 6 feet of height, and 3 feet of depth. This gives hospitals flexible options for installation. Its small size was a design requirement based on hospital's preferences (mostly economical) to have the Robot fit space constraints and permit its installation without architecture changes to the building, unlike with most other Automation solutions in the market. It occupies approximately the same space than a regular laminar flow cabinet

What are the requirements for the rooms where CytoCare™ is installed?
According to measurements done by independent laboratories, the air filtering system of the robot provides ISO 5 (class 100) conditions in the preparation chamber even if the robot is not placed in a clean room. To achieve compliance to USP 797 or to the European BMPs, the CytoCare™ unit should be placed in an ISO Class 8 (class 100,000) environment.

What kind of access does the machine need around it?
We can provide a drawing of the minimum space required by the equipment. The new machine needs to be accessible from the right side for maintenance, but as it is on wheels, it can be moved.

 